Due to frequent requests with regards to disinfection, Neuroelectrics aims to provide our customers with guidance ensuring the performance and safety characteristics of our products are not altered. 
After each use it is necessary to thoroughly clean the headcap and electrodes in order to extend their life and prevent deterioration.
In case disinfection is required after cleaning, Neuroelctrics recommends the following validated methods:
Electrodes: For disinfection, Neuroelectrics recommends using Clinell Universal Disinfectant spray. This medical device disinfectant is applied directly on all electrode surfaces. After two minutes, rinse the electrodes with water.  Once disinfected and rinsed, allow the electrodes to air dry, avoiding direct exposure to sunlight.
Headcap: Neuroelectrics recommends thorough washing with warm tap water and soap. The head cap should be washed in a manner similar to the WHO recommendations for handwashing. 

Cables and Necbox: Neuroelectrics recommends using a tissue soaked with isopropyl alcohol to carefully lay wipe all surfaces. It is not recommended to use soap and water, or any other cleaning method in these products as this could affect the electronics.